Analysis of Scene Problems for Oval Gear Flowmeter

Release time: 2026-04-16

Fault PhenomenonCauseMeasureRemarks
Oval gear does not rotate1. Debris in pipeline
2. Debris in measured liquid
3. Filter damaged
Debris enters meter, gear stuck
Disassemble and clean the meter and pipeline; repair the filter
Axial seal coupling leakagePacking worn or lack of sealing oilTighten the gland or replace the packing; add sealing oil
Pointer rotates unstablyLoose pointer or washer; moving parts not flexibleRetighten and eliminate inflexibility
Small flow error too negative (excessive negative error at low flow)Oval gear contacts measuring chamber wall due to bearing wear or chamber deformationReplace bearings; repair deformed measuring chamber and gear; ensure flexible rotation; recalibrate after repair
Large error variationLarge flow pulsation or presence of gasReduce pulsation or install a gas separatorFor 0.2-class flowmeters, the maximum/minimum error shall not exceed ±0.17%
Excessive error but not exceeding ±1%Service life exceeded or clearance changed after maintenanceRecalibrate and adjust
No signal from transmitter1. Transmitter block improperly positioned
2. Polarity reversed
Readjust position (move forward/backward/left/right)
Reconnect correctly: + to red wire, – to black wire